万通商务网
首页-产品目录

英国Harwell红色4034pmma辐射剂量计

其它参数:
产品介绍:

英国harwell 红色4034 pmma辐射剂  量计,harwell radspin丙氨酸  辐射剂  量丸,英国harwell dosimeters氨酸  辐射剂  量丸,

测量alpha射线,beta射线,gamma射线,x射线辐射!

 品名:harwell red 4034 pmma辐射剂  量计,harwell amber 3042 pmma 辐射剂  量计, harwell radspin辐射剂  量剂  ,harwell alanine pellets剂  量计。

harwell红色4034剂  量计 说明书

用于测量辐射剂  量的harwell红色4034剂  量计

1、材料:单批红色4034剂  量计由一块聚甲  基丙烯酸  甲  酯板材制作。

2、尺寸:30mm×11mm。厚度3±0.55mm

3、测量重复性:同批剂  量计,在不均匀度小于±1%的辐照场中同时辐照,用640nm波长测量,在5-50kgy剂  量范围内,其吸光率测量的变异系数小于2%。

4、校准能力:在校准范围内,所有吸光率测量数据的平均值与使用最小二乘拟合的四阶多项式的数据相比,误差在2%以内。

5、溯源性:harwell剂  量计的校准,溯源到英国国家物理实验室吸收剂  量标准。

                  与英国国家物理实验室相互比较来验证每批剂  量计的校准性能。

6、保质期:从发售日期起10年内。

7、包装:用聚酯/铝箔/聚酯/聚乙烯材料制成的带有标签  的小袋密封包装,每箱1000片。

harwell red 4034

harwell red 4034 dosimeters

red 4034 are made from radiation-sensitive poly-methylmethacrylate (pmma) in the form of optically transparent pieces individually sealed in laminate sachets. they darken when irradiated,【敏感词】the radiation-induced darkening, accurately measurable by means of a spectrophotometer, is a function of the radiation dose ABSorbed. dose range

red 4034 has a functional dose range of 5 kgy to 50 kgy which makes it idea for a wide range of irradiation processes. material

the 4034 red is made from single batches of cast sheets of poly-methylmethacrylate (pmma) dimensions

harwell dosimeters then cut the sheets down into plaques of 30mm x 11mm with a thickness of 3mm (+/- 0.55mm) packaging

after 100% inspection the dosimeters are packaged into sealed sachets made of polyester/aluminium foil/polyester/polyethylene laminate. with each sachet being labeled with a unique bar coded label.

packed 1,000 per carton.

(packing to defined aql's checked to bs6001 (iso 2859-1), single normal inspection, level ii.) measurement reproducibility

coefficient of variation of specific absorbance measurements on sets of dosimeters for the batch, simltaneously irradiated together, in a gamma radiation field uniform within +/- 1%, is 2% over the entire calibrated range, 5 to 50 kgy. calibration quality

over the calibrated range, all mean specific absorbance data points thus obtained are within 2% of a fourth-order polynomial least-squares fit to the data. shelf-life

ten years from the date of release. traceability

harwell calibration is traceable to the uk standard of absorbed dose at the national physical laboratory (npl).